Validation Report
Validation Title: Bioanalytical Method Validation for the Determination of Concentrations of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection in Cynomolgus Monkey Serum
Validation Number:
Validation Director: Zhao Xiaoping National Center for Drug Safety Evaluation and Research (NCDSER), Shanghai
Title: Bioanalytical Method Validation for the Determination of Concentrations of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection in Cynomolgus Monkey Serum Validation Director: Zhao Xiaoping Signature: _______________ Date: November 08, 2016
Except for the stages noted below, this validation was conducted in accordance with the provisions of the "Good Laboratory Practice for Non-Clinical Drug Studies" (State Food and Drug Administration Order No. 2, 2003) of the China Food and Drug Administration (CFDA). There were no adverse factors during the validation that significantly affected the quality or integrity of the study.
Certificate of Analysis (COA) for Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection (Non-GLP)
Validation Title: Bioanalytical Method Validation for the Determination of Concentrations of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection in Cynomolgus Monkey Serum
This study was inspected in accordance with the CFDA's "Good Laboratory Practice for Non-Clinical Drug Studies" (Order No. 2) and the Standard Operating Procedures (SOPs) of this institution. This validation report truthfully describes the test materials and methods and accurately reflects the raw data generated during the testing process (excluding the COA of the reference standard).
Report Validation Director Date
- Validation Protocol
- Validation Report and Records
Additionally, process-based and facility-based inspections were conducted according to the institution's SOPs.
Objective: To establish a method for determining the concentrations of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection in cynomolgus monkey serum, and to validate the method regarding standard curve and quantification range, accuracy, precision, total error of the method, specificity, selectivity, stability (partial), dilution linearity and hook effect, and parallelism.
Method: Microplates were coated with Recombinant Human ErbB2/HER2 solution at a concentration of 1 μg/mL. Detection was performed using Human IgG-heavy and light chain monkey-adsorbed Antibody solution diluted 80,000-fold. The TMB substrate catalyzed a color reaction, which was terminated with sulfuric acid solution. Absorbance was read on a microplate reader at wavelengths of 450-560 nm.
Theoretical concentration was used as the x-axis, and the difference between the measured OD value and the mean OD value of the blank replicates was used as the y-axis. Relevant parameters of the standard curve were fitted using a four-parameter logistic regression model. Sample concentrations were calculated using SoftMax Pro software versions 6.4.0 and 6.4.2.
In this validation, the quantitative range for detecting Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection or Trastuzumab for Injection was 6400–100 ng/mL. The standard curve sample concentrations were: 10,000 ng/mL (Anchor Point H), 6400 ng/mL (ULOQ), 4000 ng/mL, 2000 ng/mL, 1000 ng/mL, 500 ng/mL, 200 ng/mL, 100 ng/mL (LLOQ), 50 ng/mL (Anchor Point L), and 0 ng/mL. QC sample concentrations were LQC (250 ng/mL), MQC (1250 ng/mL), and HQC (5000 ng/mL). Samples were pre-treated with a 1:19 dilution in phosphate buffer before loading.
Results: In this validation, the correlation coefficients ($R^2$) of the fitted standard curves met the requirements. The precision (CV%) of response values among replicates at each concentration point of the standard curve was within 14.60%. The relative error (RE%) between the back-calculated values and the nominal values at each concentration point was within -12.05% to 10.95%. More than 75% of the non-zero concentration points, including at least six concentration points (including LLOQ and ULOQ), were used for the calculation of the standard curve.
Validation samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection were prepared in pooled cynomolgus monkey serum at five concentration levels. Samples were pre-treated with a 1:19 dilution in phosphate buffer before loading. The intra-assay and inter-assay accuracy results were within -18.72% to 10.60% (ULOQ and LLOQ samples were within -22.81% to 23.65%). The intra-assay and inter-assay precision results were within 0.52% to 16.32%.
The total method error (intra-assay and inter-assay) did not exceed 26.82%.
The absolute difference in inter-assay accuracy RE% for QC samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection at various concentrations was within 2.67%. The 90% confidence interval for the difference in intra-assay accuracy RE% was within -8.74% to 10.83%.
The coating antibody and detection antibody could specifically recognize the standard when the concentration of the unrelated structural analog Human IgG did not exceed 200 μg/mL.
For selectivity samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection at ULOQ and LLOQ levels, the relative deviation of at least 80% of individual samples was within ±25%. The measured values of the blank matrix were below the lower limit of quantification (LLOQ), indicating no matrix interference.
HQC and LQC samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection prepared in pooled cynomolgus monkey serum were subjected to five freeze-thaw cycles, stored at room temperature for 4 hours, stored at 2-8°C for 24 hours, pre-treated with a 1:19 dilution in phosphate buffer and stored at room temperature for 4 hours, and stored under frozen conditions (<-65°C) for 4 months and 6 months. The relative deviation between the measured concentrations and theoretical concentrations was within ±20%, indicating no significant change in concentration.
The relative deviation (RE%) between the back-calculated concentrations and theoretical concentrations of samples diluted by factors of 1000, 100, and 10 was within -5.53% to -14.43%. The coefficient of variation (CV) of the final back-calculated concentrations of the samples was within 12.70%. The maximum acceptable dilution factor for the method is 1000-fold.
The OD response values for concentrations of 100 μg/mL, 20 μg/mL, 10 μg/mL, and 5000 ng/mL decreased with decreasing concentration, indicating that no hook effect occurred.
Three real samples from the highest concentration group at the blood collection time point in toxicokinetic study 1505RD1, administered with Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection, were selected. These were diluted with blank matrix at three different dilution factors. The CV among the serially diluted samples (multiplied by the dilution factor) was within 18.39%.
The bioanalytical method validation results in this report indicate that the method meets the requirements for biological sample analysis (except for long-term stability results) and can be used for the determination of concentrations of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection in cynomolgus monkey serum.
Abbreviations/Glossary:
- Bovine Serum Albumin (BSA)
- Coefficient of Variation / Precision
- Enzyme-Linked Immunosorbent Assay (ELISA)
- High Quality Control (HQC)
- Immunoglobulin G (IgG)
- Milligram/Microgram/Nanogram
- Low Quality Control (LQC)
- Medium Quality Control (MQC)
- Quantity
- Optical Density (OD)
- Phosphate Buffered Saline (PBS)
- Relative Recovery / Accuracy
- Tetramethylbenzidine (TMB)
- Upper Limit of Quantification (ULOQ)
- Validation Title
- Validation Number
Validation Objective: The objective of this validation is to establish a method for determining the concentrations of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection in cynomolgus monkey serum, and to validate the method regarding standard curve and quantification range, accuracy, precision, total method error, specificity, selectivity, stability (partial), dilution linearity and hook effect, and parallelism, providing evidence for bioanalysis in toxicology and pharmacokinetic studies.
Addresses:
- 199 Guoshoujing Road, Zhangjiang Hi-Tech Park, Pudong, Shanghai
- 285 Copernicus Road, Zhangjiang Hi-Tech Park, Pudong, Shanghai
Validation Director: Key Personnel:
- Bioanalytical Validation: Chen Ying, Li Jinlong, Chen Jubing
Validation Schedule:
- Validation Start Date: March 18, 2016
- Experiment Start Date: March 18, 2016
- Experiment End Date (excluding partial long-term stability): September 29, 2016
- Report Completion Date: November 08, 2016
Technical Guidelines: "Guidance for Validation of Bioanalytical Quantitative Methods (Draft)", China CFDA, 2015
Reference Standards and Reagents:
| Name/Code | Code | Source | Concentration/Storage | Expiry Date/Batch No. |
|---|---|---|---|---|
| Recombinant Human ErbB2/HER2 | Stored < -65°C; 3 months at 2-8°C after thawing | Expiry for storage at -65°C: 2018.12.08 Batch: LC10JA0503, SB, Expiry: 2017.02.27 (3 months after dissolution) | ||
| Human IgG-heavy and light chain monkey-adsorbed Antibody | BETHYL | Batch: A80-319P-20, Expiry: 2017.03.10 / 2017.05.20 / 2017.06.20 | ||
| TMB Chromogenic Solution | Solarbio | Batch: 20160127, Expiry: 2019.01.27 Batch: 20160504, Expiry: 2019.05.04 | ||
| Concentrated Sulfuric Acid | Sinopharm Group Shanghai Chemical Reagent Co. | Batch: T20090115, Expiry: 2017.12.28 Batch: 20160506, Expiry: 2018.09.08 | ||
| BSA | 10735078001 | Roche | Expiry: 20160801 / 20170801 | |
| Skim Milk Powder | 4239539 | BD | Expiry: 2019.07.29 | |
| Tween 20 | Sinopharm Group Chemical Reagent Co., Ltd. | Batch: 131106, Expiry: 2016.12.02 Batch: 20160203, Expiry: 2018.04.15 | ||
| Phosphate Buffered Saline (PBS) | HYCLONE | Batch: AAF203865, Expiry: 2017.06.01 Batch: AAK208442, Expiry: 2017.10.01 | ||
| Phosphate Buffer (20×) | Sangon | Batch: BA27KA1389, Expiry: 2017.11.01 Batch: C115KA2011, Expiry: 2019.01.01 | ||
| Ultrapure Water | Self-prepared, Milli-Q Advantage A10 | |||
| Human IgG | H25345 | YEASEN | Solution | Expiry: 2017.07.02 |
Instruments and Equipment:
- Electronic Balance:
- Ultrapure Water System:
- Refrigerator/Freezer: EQ2015B0, Haier DW-86L388A
- Microplate Reader:
- Plate Washer:
- Vortex Mixer:
- Biochemical Incubator:
- Magnetic Stirrer:
- Microplate Shaker:
- Electronic Pipette:
Safety Precautions: Laboratory protective clothing, gloves, and other protective equipment must be worn during the preparation of reference standards and sample handling.
Return of Unused Reference Standards: Unused reference standards remaining after the completion of this validation shall be returned to the sponsor by the Analysis and Bioanalysis Department II after all tests are completed, or disposed of upon written authorization from the sponsor.
Solution Preparation: The preparation of relevant reagents and samples listed in this report is for clarifying the preparation method only. Specific preparation volumes should be adjusted proportionally according to actual conditions and recorded in the original records.
Preparation of Recombinant Human ErbB2/HER2 Coating Solution: Dilute the Recombinant Human ErbB2/HER2 coating antibody with phosphate buffer to a concentration of 1 μg/mL. Prepare fresh for use.
Preparation of Wash Buffer: Add 20× PBS to ultrapure water to prepare 1× PBS solution and mix well. Add Tween 20 to the prepared 1× PBS solution, mix well, to prepare 0.05% PBS-Tween wash buffer. Store at room temperature; valid for 3 months.
Preparation of Blocking Solution: Weigh an appropriate amount of skim milk powder and dilute with phosphate buffer to prepare the skim milk solution. Prepare fresh for use.
Preparation of 1% BSA: Accurately weigh 1 g of BSA powder, add to 100 mL of phosphate buffer, and stir with a magnetic stirrer until completely dissolved. Valid for 3 months when stored at 2-8°C.
Preparation of Secondary Antibody Solution: Dilute the secondary antibody (Human IgG-heavy and light chain monkey-adsorbed Antibody) 80,000-fold with 1% BSA solution. Prepare fresh for use.
Preparation of Stop Solution: Slowly add 10.87 mL of concentrated sulfuric acid to 100 mL of ultrapure water while stirring. After cooling to room temperature, transfer to a volumetric flask. Store at room temperature; valid for 3 months.
Pooled Cynomolgus Monkey Serum: Mix blank serum from 35 individual cynomolgus monkeys and store frozen at -20°C. Valid for 2 years (calculated based on the earliest expiry date of the individual sera included in the pool).
Experimental Procedure: The following experiments were conducted at room temperature unless incubation conditions are specified.
- Coating: Add 100 μL of 1 μg/mL Recombinant Human ErbB2/HER2 coating solution to each well of a 96-well plate. Seal the plate with a sealing film and incubate overnight (14-18 hours) in a 2-8°C refrigerator.
- Washing: Discard the liquid from the wells. Add 300 μL of wash buffer to each well and wash three times. Blot dry on paper towels.
- Blocking: Add 300 μL of blocking solution to each well. Seal the plate with a sealing film and incubate at 25°C for at least 2 hours.
- Sample Addition: Add 100 μL of standard curve samples, QC samples, and unknown samples (if any) to the respective wells. Seal the plate with a sealing film and incubate at 25°C for 1 h ± 3 min.
- Secondary Antibody Addition: Add 100 μL of secondary antibody solution to each well. Seal the plate with a sealing film and incubate at 25°C for 1 h ± 3 min.
- Washing: Discard the liquid from the wells. Add 300 μL of wash buffer to each well and wash four times. Blot dry on paper towels.
- Color Development: Add 100 μL of TMB chromogenic solution to each well. Incubate at 25°C in the dark for 15 min ± 5 min.
- Termination: Add 100 μL of 2M H₂SO₄ solution to each well to terminate the reaction.
- Reading: Read the plate using a microplate reader at wavelengths of 450-560 nm.
Standard Curve and Quantification Range: A series of working solutions of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection at different concentrations were prepared in pooled cynomolgus monkey serum. The concentrations were: 10,000 ng/mL (Anchor Point H), 6400 ng/mL (ULOQ), 4000 ng/mL, 2000 ng/mL, 1000 ng/mL, 500 ng/mL, 200 ng/mL, 100 ng/mL (LLOQ), 50 ng/mL (Anchor Point L), and 0 ng/mL (blank sample containing cynomolgus monkey blank serum without the antibody). Standard curve samples were prepared by pre-treating with a 1:19 dilution in phosphate buffer before loading.
Theoretical standard concentration was used as the x-axis, and the difference between the mean OD value of standard replicates and the mean OD value of blank replicates was used as the y-axis. A four-parameter logistic regression model was used to fit the curve and provide relevant parameters for the standard curve.
During validation, at least 6 independent analytical batches on different dates must be measured. For each standard curve, the correlation coefficient $R^2$ must meet requirements. In addition to the effective standard curve concentration gradients, anchor points can be used to assist in curve fitting. Concentration points with errors due to known or unknown causes may be excluded during fitting.
The precision (CV%) of concentrations among replicates at each point of the standard curve must be within 20% (CV% for LLOQ and ULOQ within 25%). The relative deviation between the back-calculated value and the nominal value at each concentration point must be within ±20% (deviation for LLOQ and ULOQ within ±25%).
More than 75% of the non-zero concentration points, including at least six concentration points (including LLOQ and ULOQ), must be used for the calculation of the standard curve and must meet the above criteria.
Accuracy, Precision, and Total Method Error: Validation samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection were prepared in pooled cynomolgus monkey serum at five concentrations: LLOQ (100 ng/mL), LQC (250 ng/mL), MQC (1250 ng/mL), HQC (5000 ng/mL), and ULOQ (6400 ng/mL). Samples were pre-treated with a 1:19 dilution in phosphate buffer before loading to evaluate intra-assay and inter-assay accuracy, precision, and total method error.
For intra-assay testing, three sets of QC samples (each set in replicate) were prepared for each concentration. The sample concentration is the mean value.
For inter-assay testing, at least 6 independent analytical batches were measured on different dates, using different instruments and personnel.
Accuracy is expressed as relative error (RE%), and results should be within ±20% (ULOQ and LLOQ samples within ±25%). Precision is expressed as coefficient of variation (CV%), and results should be within 20% (ULOQ and LLOQ samples within 25%). Total method error (sum of the absolute value of % relative deviation and % coefficient of variation) should not exceed 30% (ULOQ and LLOQ samples not exceeding 40%).
At least 67% of the QC samples must meet the above criteria, with at least two QC samples meeting the criteria at each concentration level. Additionally, the mean value of QC samples at each concentration must meet the above criteria.
The absolute difference in inter-assay accuracy RE% for QC samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection at various concentrations should be within 20% (ULOQ and LLOQ samples within 25%).
The 90% confidence interval for the difference in accuracy RE% of QC samples of the same concentration of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection within each analytical batch must be within ±30% (ULOQ and LLOQ samples within ±35%).
Specificity: Samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection at LLOQ and ULOQ levels (3 replicates, mean result taken) and blank matrix without the antibodies were prepared in pooled cynomolgus monkey serum. Increasing concentrations of Human IgG samples were added.
It is required that at least 80% of the QC samples at the same concentration have a relative deviation within ±25%, and the measured value of the blank matrix without the antibodies should be below the lower limit of quantification.
The concentration of Human IgG samples added when specificity validation passes must be noted in the validation report.
Selectivity: Selectivity samples were prepared by adding Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection at LLOQ and ULOQ levels to serum from at least 10 individual cynomolgus monkeys. Individual blank matrices were also included and measured alongside the standard curve. Three replicates were prepared for selectivity samples at each concentration, and the mean result was taken.
The acceptance criterion for results is that the relative deviation of at least [missing number in source, implied 80% based on context] of the selectivity samples at the same concentration is within the range. Additionally, the measured value of the blank matrix without the antibodies should be below the lower limit of quantification.
Stability: HQC and LQC samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection were prepared in pooled cynomolgus monkey serum. The concentration of stability QC samples was calculated using a freshly prepared standard curve. The relative deviation (RE%) from the theoretical concentration must be within ±20%.
Stability of the analyte in the experimental matrix includes:
- Freeze-Thaw Stability: Three sets of HQC and LQC concentration samples were prepared in pooled cynomolgus monkey serum. They were frozen (<-65°C) for at least 12 hours and then thawed at room temperature. This cycle was repeated five times. The results are the average of the three sets of QCs.
- Short-Term Stability: Six sets of HQC and LQC concentration samples were prepared in pooled cynomolgus monkey serum. Three sets were placed at room temperature for 4 hours, and three sets were placed at 2-8°C for 24 hours before analysis.
- Long-Term Stability: Nine sets of HQC and LQC concentration samples were prepared in pooled cynomolgus monkey serum. They were stored under frozen conditions (<-65°C) for 4 months, 6 months, and 8 months. Three sets were analyzed at each time point, and the average result was taken.
- Bench-Top Stability: Three sets of HQC and LQC concentration samples were prepared in pooled cynomolgus monkey serum. They were pre-treated with a 1:19 dilution in phosphate buffer and placed at room temperature for 4 hours. The results are the average of the three sets of QCs.
Dilution Linearity and Hook Effect: An appropriate amount of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection was added to pooled cynomolgus monkey serum to prepare samples with concentrations higher than the ULOQ. These were diluted with pooled cynomolgus monkey serum by factors of 1:1000, 1:100, and 1:10 to fall within the standard curve concentration range. At least 3 replicates were tested for each dilution factor, and the mean back-calculated concentration was taken.
The relative deviation (RE%) between the back-calculated concentration and the theoretical concentration for samples at each dilution factor should be within ±20%, and the coefficient of variation (CV) of the final back-calculated concentrations of all samples should be within 20%.
The same operation and acceptance criteria were applied to Trastuzumab for Injection.
To assess the hook effect, the OD response values of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection samples at concentrations of 100 μg/mL, 20 μg/mL, 10 μg/mL, and 5000 ng/mL were examined. At least three replicates were tested for each concentration point, and the mean OD was taken. If the response value decreases with decreasing concentration, it indicates that no hook effect occurred.
Parallelism: Three real samples from the highest concentration group near the blood collection time point, administered with Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection, were selected. They were diluted with blank matrix to at least 3 different concentrations (diluted concentrations within the quantification range of the standard curve) and measured. Three replicates were prepared for each sample at each dilution factor.
The coefficient of variation (CV%) among the serially diluted samples (multiplied by the dilution factor) was within 30%.
Data Analysis: All measurement results were collected using SoftMax Pro software versions 6.4.0 and 6.4.2. Data were calculated using SoftMax Pro software versions 6.4.0 and 6.4.2 and Microsoft Excel 2007. Statistical analysis was performed using IBM SPSS Statistics 21.
Sample concentration results are retained to one decimal place. Mean, CV%, RE%, and total error results are retained to two decimal places. OD values are retained to three decimal places.
Due to differences in significant figure rounding methods among different programs, numerical values in some tables may differ slightly from individually calculated results, but this does not affect the integrity and continuity of the data.