Discussion
Hyperglycemia is a metabolic disorder. With the increasing maternal age of childbearing women in China, the incidence of pregnancy complicated by hyperglycemia shows a gradual upward trend. Patients with pregnancy complicated by hyperglycemia often present with other metabolic abnormalities, such as hyperinsulinemia and insulin resistance. These conditions exacerbate the adverse effects on the intrauterine environment and significantly increase the risks of gestational hypertension, preeclampsia, and preterm birth. Furthermore, pregnancy complicated by hyperglycemia increases the incidence of complications such as fetal growth restriction, macrosomia, and amniotic fluid abnormalities, thereby increasing the difficulty of delivery. Hyperglycemia not only elevates the risk of short-term and long-term adverse outcomes for pregnant women but also affects the metabolic outcomes of offspring to a certain extent. Standardized management of blood glucose during pregnancy can reduce maternal and fetal complications associated with hyperglycemia in both the short and long term, becoming a key link in the prevention and control of diabetes under the concept of full life-cycle health.
Studies in China have found that patients using Continuous Glucose Monitoring (CGM) systems exhibit significantly reduced glycated hemoglobin (HbA1c) levels, as well as significantly decreased duration and frequency of hypoglycemic events. Meanwhile, CGM systems significantly improve patient experience. Abroad, CGM technology has been applied to patients with pregnancy complicated by type 1 diabetes, proving that CGM helps achieve HbA1c targets and reduces the risks of macrosomia and neonatal hypoglycemia in pregnant women with type 1 diabetes. In China, there is clear encouragement for the use and gradual standardization of new blood glucose monitoring technologies, including minimally invasive, non-invasive, and remote monitoring, during pregnancy.
In the traditional concept of blood glucose monitoring, capillary blood glucose meters are the most widely accepted due to their ease of use. They represent the longest-standing and most fundamental method of self-monitoring of blood glucose (SMBG). However, their accuracy depends on the correct use of the device; improper operation may affect the accuracy of blood glucose measurements. Additionally, SMBG is limited by the inherent constraints of blood glucose monitoring and the patient's ability to interpret results. Finger-stick blood sampling may cause discomfort to patients. Insufficient monitoring frequency leads to a lack of sufficient evidence for judging mean blood glucose, glucose variability, or the incidence of hypoglycemia. Conversely, overly frequent monitoring may cause anxiety in some patients. This is particularly true for pregnant women, who may experience greater psychological pressure, and emotional fluctuations can negatively impact blood glucose levels.
Compared with traditional SMBG, CGM not only provides a comprehensive and detailed understanding of blood glucose status, forming dynamic glucose profile reports for systematic analysis, but also allows for continuous assessment of the efficacy and safety of treatment plans while formulating individualized therapeutic strategies. However, unlike capillary blood glucose monitoring, CGM technology monitors glucose concentration in subcutaneous interstitial fluid via glucose sensors. There are discrepancies between CGM-measured glucose levels and those measured by blood glucose meters, which is currently the main reason why CGM technology has not been fully popularized.
In this study, when discrepancies occurred between CGM and SMBG readings, the SMBG results were taken as the standard. When patients experienced hypoglycemia or abnormally elevated blood glucose, SMBG was performed to calibrate using capillary blood glucose. Multiple comparisons of blood glucose levels revealed that the difference between the CGM system and SMBG was 1–2 mmol/L, indicating no significant discrepancy.