我是梅斯医学智能翻译助理,请提供需要翻译的内容
Major Software Systems
The primary computerized systems used in this validation include:
| Software Name | Version Number | Purpose |
|---|---|---|
| [Software Name Missing in Source] | 6.4.0 and 6.4.2 | Data acquisition, statistics, and analysis |
| [Software Name Missing in Source] | [Version Missing] | Data entry, statistics, and analysis |
| [Software Name Missing in Source] | [Version Missing] | Statistical data analysis |
Description of Analysis Batches
| Analysis Batch ID | Experiment Date | Experiment Content | Result Acceptance |
|---|---|---|---|
| [ID Missing] | [Date Missing] | Selectivity of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection (Individuals 1-7) | No* |
| [ID Missing] | [Date Missing] | Selectivity of Trastuzumab for Injection (Individuals 1-7) | No* |
| [ID Missing] | [Date Missing] | Selectivity of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection (Individuals 8-10) | No* |
| [ID Missing] | [Date Missing] | Specificity of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection | No* |
| [ID Missing] | [Date Missing] | Dilution linearity, hook effect, stability after pre-treatment | No* |
| [ID Missing] | [Date Missing] | Stability after three freeze-thaw cycles, at room temperature, and at 2–8°C | No* |
| [ID Missing] | [Date Missing] | Stability after four and five freeze-thaw cycles | No* |
| [ID Missing] | [Date Missing] | Long-term stability | No* |
*: The analysis batch was deemed failed because precision and accuracy did not meet the acceptance criteria.
Eight independent analysis batches were conducted by three analysts over four days. The results of the analysis batches are presented in [Table Reference]. The correlation coefficient ($R^2$) of each fitted standard curve met the requirements. The precision (CV%) of the response values among replicate wells at each concentration point of the standard curve ranged from 0.10% to 19.10%. The relative error (RE%) between the back-calculated values and the nominal values at each concentration point ranged from -12.05% to 12.45%. More than 75% of the non-zero concentration points, with at least six concentration points (including LLOQ and ULOQ), were used for the calculation of the standard curve.
Table: Standard Curve Results
| Theoretical Concentration of Standard Curve (ng/mL) | ... |
|---|
Validation samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection were prepared in pooled cynomolgus monkey serum at five concentrations. For intra-batch assessment, three sets of QC samples (each set in duplicate) were prepared for each concentration. The sample concentration was expressed as the mean. For inter-batch assessment, eight independent analysis batches were performed on two different models of microplate readers by different personnel over four working days. Two of these analysis batches were deemed failed due to insufficient QC samples meeting the criteria and were excluded from the intra-batch and inter-batch result calculations.
The results for intra-batch and inter-batch accuracy, precision, and total method error are shown in [Table References].
The intra-batch and inter-batch accuracy results for QC samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection at various concentrations ranged from -18.72% to 10.60% (for ULOQ and LLOQ samples, the range was -22.81% to 23.65%). The intra-batch and inter-batch precision results ranged from 0.52% to 16.32%.
At least 67% of the QC samples met the acceptance criteria, with at least two QC samples meeting the criteria at each concentration level. Additionally, the mean values of the QC samples at each concentration met the aforementioned standards.
The absolute difference in inter-batch accuracy RE% for QC samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection at various concentrations was within 2.67%.
The 90% confidence interval of the difference in accuracy RE% for QC samples of the same concentration of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection within each analysis batch ranged from -8.74% to 10.83%, as shown in [Table Reference].
Table: Intra-batch and Inter-batch Results for Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection
*: Excluded from intra-batch and inter-batch data calculations because the RE% did not meet the acceptance criteria.
Table: Intra-batch and Inter-batch Results for Trastuzumab for Injection
BQL: Below the Quantitation Limit (100 ng/mL); excluded from intra-batch and inter-batch data calculations.
Table: Comparison of Intra-batch and Inter-batch Accuracy RE% for Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection
| Parameter | Value |
|---|---|
| Difference in Intra-batch RE (%) | ... |
| Upper Limit of 90% Confidence Interval | ... |
| Lower Limit of 90% Confidence Interval | ... |
| Absolute Difference in Inter-batch RE (%) | ... |
Samples of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection at LLOQ and ULOQ levels (prepared in pooled cynomolgus monkey serum, n=3 replicates, mean result) and blank matrix (without the addition of the antibodies) were spiked with increasing concentrations of Human IgG (50 μg/mL, 100 μg/mL, 200 μg/mL). The results are shown in [Table Reference].
The results indicated that the relative deviation of ULOQ samples for both Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection was within ±25%. For LLOQ samples, at least 80% (5/6) showed relative deviation within ±25%. The measured values of the blank matrix were below the lower limit of quantification (denoted as BQL).
These results demonstrate that the coating antibody and detection antibody can specifically recognize the standard in the presence of the unrelated structural analog Human IgG at concentrations up to 200 μg/mL.
Table: Specificity Results
| Theoretical Concentration | Specificity Concentration (μg/mL) | Mean Concentration |
|---|---|---|
| ... | ... | ... |
BQL: Below the Quantitation Limit (100 ng/mL).
Selectivity samples were prepared by adding Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection at LLOQ and ULOQ levels to serum from 10 individual cynomolgus monkeys. Individual blank matrices were included and analyzed alongside the standard curve. The results are shown in [Table References].
The results showed that for Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection, the relative deviation of selectivity samples at the ULOQ level was within ±25% for 9/10 (90%) individuals, and at the LLOQ level for 10/10 (100%) individuals. For Trastuzumab for Injection, the relative deviation of selectivity samples at both ULOQ and LLOQ levels was within ±25% for 10/10 (100%) individuals. The measured values of the blank matrix (without the addition of the antibodies) were below the lower limit of quantification (BQL). No matrix interference was observed, indicating that the method has good selectivity for the tested monkey serum.
Table: Selectivity Results for Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection
| Selective Individual | Mean Back-calculated Concentration |
|---|---|
| Individual 1 | ... |
Table: Selectivity Results for Trastuzumab for Injection
Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection was prepared in pooled cynomolgus monkey serum as HQC and LQC samples. These samples were subjected to five freeze-thaw cycles, stored at room temperature for 4 hours, stored at 2–8°C for 24 hours, pre-treated with a 1:19 dilution in phosphate-buffered saline (PBS) and stored at room temperature for 4 hours, or stored under frozen conditions (<-65°C) for 4 months and 6 months. The relative deviation between the measured concentrations and the theoretical concentrations was within ±20% in all cases, with no significant change in concentration observed. The results are shown in [Table Reference].
The stability results for QC samples stored under frozen conditions (<-65°C) for 8 months will be presented in a report amendment.
Table: Stability Results
| Concentration (ng/mL) | 1 Freeze-Thaw Cycle | 2 Freeze-Thaw Cycles | 3 Freeze-Thaw Cycles | 4 Freeze-Thaw Cycles | 5 Freeze-Thaw Cycles | Room Temp 4h | RT 4h after 1:19 PBS Pre-treatment | Frozen 4 Months |
|---|---|---|---|---|---|---|---|---|
| ... | ... | ... | ... | ... | ... | ... | ... | ... |
Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection was added to pooled cynomolgus monkey serum to prepare samples with concentrations higher than the ULOQ (1 mg/mL, 100 μg/mL, 10 μg/mL). These were diluted 1:1000, 1:100, and 1:10 to fall within the standard curve concentration range. Three replicates were prepared for each dilution factor, and the mean back-calculated concentration was determined. The same procedure was applied to Trastuzumab for Injection. The results are shown in [Table Reference].
The results showed that the relative error (RE%) between the back-calculated concentrations and the theoretical concentrations for samples at each dilution factor was within ±20% (ranging from -5.53% to -14.43%). Furthermore, the coefficient of variation (CV%) for the final back-calculated concentrations of all samples was within 20% (not exceeding 12.70%). The maximum acceptable dilution factor for the method is 1000-fold.
The hook effect results are shown in [Table Reference]. The OD response values of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection samples at concentrations of 100 μg/mL, 20 μg/mL, 10 μg/mL, and 5000 ng/mL were evaluated. The results showed that the response values decreased as the concentration decreased, indicating the absence of a hook effect.
Table: Dilution Linearity Results
| Prepared Concentration (ng/mL) | Back-calculated Concentration |
|---|---|
| ... | ... |
Table: Hook Effect Results
| Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection | |
|---|---|
| Concentration (ng/mL) | Mean OD |
| ... | ... |
Three real samples from the highest concentration group at the blood collection time point in the toxicokinetic study (Study No. 1505RD1) administered with Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection were selected. These samples were diluted with blank matrix to three different concentrations (within the quantitative range of the standard curve). Three replicates were prepared for each sample at each dilution factor. The measurement results are shown in [Table Reference].
The coefficient of variation (CV%) among the samples, after multiplying the serially diluted samples by their respective dilution factors, was within 30% (not exceeding 18.39%).
Table: Parallelism Results
| Sample ID | Back-calculated Concentration (ng/mL) | Sample Concentration (μg/mL) |
|---|---|---|
| ... | ... | ... |
The bioanalytical method validation results in this report (excluding partial long-term stability results) indicate that the method meets the requirements for biological sample analysis and can be used for the determination of Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection and Trastuzumab for Injection concentrations in cynomolgus monkey serum.
Documents related to this validation, including the validation protocol and amendments, raw records, and validation reports, will be archived in the institution's archive room until five years after the drug is marketed.
Appendix I: Validation Protocol and Protocol Amendments
Bioanalytical Method Validation for the Determination of RE1505-A and RE1505-B Concentrations in Cynomolgus Monkey Serum
Sponsor Representative: Zhang Chengju Head of Quality Assurance Department: Ma Jing
Objective: The purpose of this validation is to establish a method for determining the concentrations of RE1505-A and RE1505-B in cynomolgus monkey serum and to validate the method regarding standard curve and quantitative range, accuracy, precision, total method error, specificity, selectivity, stability (partial), dilution linearity and hook effect, and parallelism, providing evidence for bioanalysis in toxicology and pharmacokinetic studies.
Email Address: [Missing] Bioanalysis Team: Chen Ying, Li Jinlong, Chen Jubing
Proposed Validation Schedule:
- Validation Protocol Approval Date: [Date Missing]
- Experiment Start Date: [Date Missing]
- Experiment End Date: [Date Missing]
- Proposed Draft Report Completion Time: [Time Missing]
GLP Compliance: Except for the Certificate of Analysis (COA) of the standard RE1505-A (completed by the sponsor under non-GLP conditions), all stages of this validation were completed at our institution. The study complied with this validation protocol, the institution's relevant SOPs, and the following regulatory guidelines:
- China CFDA, Good Laboratory Practice for Non-Clinical Drug Studies (Order No. 2, 2003).
- Guiding Principles for the Validation of Bioanalytical Methods for Quantitative Analysis of Biological Samples (Draft), China CFDA, 2015.
Storage Conditions:
- Below -65°C; 3 months at 2–8°C after thawing.
- Expiry date for storage at -65°C is 2018.12.08. Post-thaw expiry dates will be recorded in the raw records and report.
- Below -65°C; 3 months at 2–8°C after thawing.
Reagents and Materials:
- Recombinant Human ErbB2/HER2: Batch number/source/expiry date recorded in raw records and summary report.
- Human IgG-heavy and light chain monkey-adsorbed Antibody: Batch number/source/expiry date recorded in raw records and summary report.
- TMB Single-Component Substrate Solution: Batch number/source/expiry date recorded in raw records and summary report.
- Concentrated Sulfuric Acid: Batch number/source/expiry date recorded in raw records and summary report.
- Skim Milk Powder: [Details Missing]
- Tween 20: [Details Missing]
- Phosphate Buffered Saline (PBS): [Details Missing]
- Phosphate Buffer (20×): [Details Missing]
- Ultrapure Water: [Details Missing] (If other reagents are used, specific information will be reflected in the raw records and report.)
Instrument Model/Equipment ID: Recorded in raw records and summary report. (If other instruments are used, specific information will be reflected in the raw records and report.)
The preparation of relevant reagents and samples listed in this protocol is intended to clarify the preparation method. Specific preparation volumes may be adjusted proportionally based on actual conditions.
Preparation Procedures:
- Dilute Recombinant Human ErbB2/HER2 coating antibody with phosphate buffer to a concentration of 1 μg/mL. Prepare fresh before use.
- Weigh an appropriate amount of skim milk powder and dilute with phosphate buffer to prepare a 5% skim milk solution.
- Preparation of 1% BSA.
- Preparation of 2M H₂SO₄ solution.
- Pool blank serum from at least 20 individual cynomolgus monkeys and store frozen at -20°C. Valid for 2 years.
Assay Procedure:
- Add 100 μL of 1 μg/mL Recombinant Human ErbB2/HER2 coating solution to each well of a 96-well plate. Seal the plate with a sealing film and incubate overnight (14–18 hours) in a 2–8°C refrigerator.
- Discard the liquid from the wells. Wash the plate three times with 300 μL of wash buffer per well. Blot dry on paper towels.
- Add 300 μL of blocking solution to each well. Seal the plate with a sealing film and incubate at 25°C for at least 2 hours. (If not used on the same day, store at 2–8°C and use within three days. Sealed plates removed from 2–8°C storage must be equilibrated at room temperature for at least 30 minutes before use.)
- Add 100 μL of standard curve samples, QC samples, and unknown samples (if any) to the respective wells. Seal the plate with a sealing film and incubate at 25°C for 1 h ± 3 min.
- Add 100 μL of secondary antibody solution to each well. Seal the plate with a sealing film and incubate at 25°C for 1 h ± 3 min.
- Discard the liquid from the wells. Wash the plate four times with 300 μL of wash buffer per well. Blot dry on paper towels.
- Add 100 μL of TMB substrate solution to each well. Incubate at 25°C in the dark for 15 min ± 5 min.
- Stop reaction: [Stop Solution Details Missing in Source, implied by context].
- Read the plate using a microplate reader at wavelengths of 450–560 nm.
Standard Curve Preparation: Prepare a series of standard curve working solutions of RE1505-A in pooled cynomolgus monkey serum at the following concentrations: 10,000 ng/mL (Anchor Point H), 6,400 ng/mL (ULOQ), 4,000 ng/mL, 2,000 ng/mL, 1,000 ng/mL, 500 ng/mL, 200 ng/mL, 100 ng/mL (LLOQ), 50 ng/mL (Anchor Point L), and 0 ng/mL (blank sample consisting of cynomolgus monkey blank serum without RE1505-A). Prior to loading, pre-treat with a 1:19 dilution in phosphate buffer to obtain the standard curve samples.
Acceptance Criteria for Standard Curve:
- The precision (CV%) among replicate wells at each concentration point of the standard curve must be within 20% (within 25% for LLOQ and ULOQ).
- The relative error (RE%) between the back-calculated values and the nominal values at each concentration point must be within ±20% (within ±25% for LLOQ and ULOQ).
- More than 75% of the non-zero concentration points, with at least six concentration points (including LLOQ and ULOQ), must be used for the calculation of the standard curve and meet the above criteria.
Accuracy and Precision Validation: Prepare validation samples of RE1505-A and RE1505-B in pooled cynomolgus monkey serum at five concentrations: LLOQ (100 ng/mL), LQC (250 ng/mL), MQC (1,250 ng/mL), HQC (5,000 ng/mL), and ULOQ (6,400 ng/mL). Pre-treat with a 1:19 dilution in phosphate buffer before loading to evaluate intra-batch and inter-batch accuracy, precision, and total method error.
- The absolute difference in inter-batch accuracy RE% for QC samples of RE1505-A and RE1505-B at various concentrations should be within 20% (within 25% for ULOQ and LLOQ samples).
- The 90% confidence interval of the difference in accuracy RE% for QC samples of the same concentration of RE1506-A and RE1506-B [Note: Source says RE1506, likely typo for RE1505] within each analysis batch must be within ±30% (within ±35% for ULOQ and LLOQ samples).
Specificity: Prepare RE1505-A and RE1505-B samples at LLOQ and ULOQ levels in pooled cynomolgus monkey serum (n=3 replicates, mean result) and blank matrix (without RE1505-A or RE1505-B). Spike with increasing concentrations of Human IgG samples.
- Acceptance Criterion: At least 80% of QC samples at the same concentration must have a relative deviation within ±25%. The measured value of the blank matrix without RE1505-A or RE1505-B must be below the lower limit of quantification.
Selectivity: Add RE1505-A and RE1505-B at LLOQ and ULOQ levels to serum from at least 10 individual cynomolgus monkeys to prepare selectivity samples. Include individual blank matrices and analyze alongside the standard curve. Prepare three replicates for each concentration of selectivity samples and take the mean.
- Acceptance Criterion: At least 80% of the selectivity samples at the same concentration must have a relative deviation within ±25%. The measured value of the blank matrix without RE1505-A or RE1505-B must be below the lower limit of quantification.
Stability: Prepare RE1505-A HQC and LQC samples in pooled cynomolgus monkey serum.
- Freeze-Thaw Stability: Prepare three sets of HQC and LQC samples. Freeze at <-65°C for at least 12 hours, then thaw at room temperature. Perform three freeze-thaw cycles. Take the mean of the three QC sets.
- Short-term Temperature Stability: Prepare six sets of HQC and LQC samples. Store three sets at room temperature for 4 hours and three sets at 2–8°C for 24 hours before analysis.
- Processed Sample Stability: Prepare three sets of HQC and LQC samples each. Pre-treat with a 1:19 dilution in phosphate buffer, then store at room temperature for 4 hours. Take the mean of the three QC sets.
Dilution Linearity: Add an appropriate amount of RE1505-A to pooled cynomolgus monkey serum to prepare samples with concentrations higher than the ULOQ. Dilute 1:1000, 1:100, and 1:10 with pooled cynomolgus monkey serum to fall within the standard curve concentration range. Prepare at least three replicates for each dilution factor and take the mean back-calculated concentration.
- Perform the same operation for RE1505-B and apply the same acceptance criteria.
Hook Effect: Evaluate the OD response values of RE1505-A and RE1505-B samples at concentrations of 100 μg/mL, 20 μg/mL, 10 μg/mL, and 5,000 ng/mL. Prepare at least three replicates for each concentration point and take the mean OD value.
Parallelism: Select three real samples from near the blood collection time point of the highest concentration group administered with RE1505-A. Dilute with blank matrix to at least three different concentrations (within the quantitative range of the standard curve) and measure. Prepare three replicates for each sample at each dilution factor.
- Acceptance Criterion: The coefficient of variation (CV%) among the samples, after multiplying the serially diluted samples by their respective dilution factors, must be within 30%.
Data Processing: All measurement results were acquired using SoftMax Pro software. Data were calculated using SoftMax Pro software and Microsoft Excel 2007. Statistical analysis was performed using IBM SPSS Statistics 21.
- Sample concentration results are retained to one decimal place.
- Mean, CV%, RE%, and Total Error results are retained to two decimal places.
- OD values are retained to three decimal places.
$$CV% = \frac{\text{Standard Deviation}}{\text{Mean}} \times 100%$$
Major Computerized Systems Used in This Validation Include (but are not limited to):
- [Software Name Missing]: Version 6.4.0 or 6.4.2
Report: Upon completion of the validation (excluding partial long-term stability results), the validation lead will complete the report. Any deviations that may affect the quality and integrity of the study will be attached to the final report. Partial long-term stability validation results will be presented in the form of a report amendment. Documents related to this study, including the validation protocol and amendments (if any), raw records, and validation report, will be archived in the institution's archive room until five years after the drug is marketed.
Validation Protocol Amendment - 001
Amendment Item No.: 001
Original Protocol:
- Prepare three sets of HQC and LQC samples in pooled cynomolgus monkey serum. Freeze at <-65°C for at least 12 hours, then thaw at room temperature. Evaluate freeze-thaw stability for three cycles. Take the mean of the three QC sets.
- Prepare nine sets of HQC and LQC samples in pooled cynomolgus monkey serum. Store under frozen conditions (<-65°C) for 1 month, 3 months, and 6 months. Analyze three sets of samples at each time point. Take the mean of the results.
Amended Protocol:
- Prepare three sets of HQC and LQC samples in pooled cynomolgus monkey serum. Freeze at <-65°C for at least 12 hours, then thaw at room temperature. Evaluate freeze-thaw stability for five cycles. Take the mean of the three QC sets.
Reason for Amendment: To accommodate the time interval between the collection of real biological samples and analysis, as well as the actual number of freeze-thaw cycles experienced by the samples, the duration of the long-term stability project was extended, and the number of freeze-thaw stability validation cycles was increased.
Validation Protocol Amendment Approval: Chang Yan
Appendix II: Standard COA
Certificate of Analysis for Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection
Page: 1/1
| Item | Specification / Requirement | Result |
|---|---|---|
| Name | Recombinant Humanized Anti-HER2 Monoclonal Antibody for Injection | |
| Batch No. | S20151001 | |
| Specification | 150 mg/vial (reconstituted with 10 mL) | |
| Dosage Form | Lyophilized preparation | |
| Manufacturing Date | 2015.10.30 | |
| Expiry Date | 2018.10.29 | |
| Appearance | Should be a white or off-white loose mass, with no signs of melting. | Complies |
| Clarity after Reconstitution | Should be a colorless or pale yellow clear liquid, may exhibit opalescence, should not be turbid. | Complies |
| Visible Particles | Should comply with regulations. | Complies |
| Fill Volume Variation | Should comply with regulations. | Complies |
| Reconstitution Time | Add sterile water for injection at 20–25°C as indicated. Shake gently. Should dissolve completely within 15 minutes. | Complies |
| Water Content | Should comply with regulations. | Complies |
| Insoluble Particles | Particles ≥10 μm: ≤6000/vial | Complies |
| Osmolality | Should comply with regulations. | Complies |
| Protein Content (UV) | Should be 90%–110% of the labeled amount. | Complies |
| Isoelectric Point (pI) | pI: 8.5–9.0, Main peak difference ≤ ±0.3 pI | pI: 8.6, Main peak difference: 0 |
| Reduced CE-SDS | Non-glycosylated heavy chain ≤ 2% | Complies |
| Non-reduced CE-SDS | Total small molecule impurities ≤ 8% Aggregates ≤ 5% | Complies |
| Charge Variants (CE-SDS/iCIEF) | Acidic peak: ≤ 35% Basic peak: ≤ 15% | Acidic peak: 26% Basic peak: 8% |
| Tween 20 Content | Should comply with regulations. | Complies |
| Biological Activity (Proliferation Inhibition) | 50–150% of the reference product. | Complies |
| Bacterial Endotoxin | < 0.1 EU/mg protein | Complies |
| Sterility | No growth. | Complies |
| Abnormal Toxicity | Complies with Pharmacopoeia requirements. | Complies |
Conclusion: The test results comply with the standards. Qualified.
Unit (Official Seal): [Seal Missing]
Appendix III: Validation Protocol Deviations
Deviation Content: The protocol stated: "The preparation method for 2M H₂SO₄ solution is to slowly add 10.87 mL of concentrated sulfuric acid to 100 mL of ultrapure water, stirring while adding, and transferring to a reagent bottle after cooling to room temperature."
- The actual raw records do not reflect the operation of "stirring while adding, and transferring to a reagent bottle after cooling to room temperature."
- The concentration of the H₂SO₄ solution prepared according to this method was not 2M.
Impact on Validation Study Quality and Integrity: The quality and integrity of the validation were not affected.